Specifications
|
For a successful design the following fundamental requirements must be achieved while considerable effort should be given to obtaining the secondary requirements.
Fundamental
Requirements
�
Measure
and catalog current
draw of the roughing pump.
�
Measure
and catalog temperature
of the roughing pump.
�
Measure and catalog oil level of the roughing pump.
�
Monitoring with a minimum sampling rate of 1sample/min.
�
Design
Budget $500.
�
Unit
Budget $200.
�
Efficiently
designed for 1-12
roughing pumps
Secondary
Requirements
�
Measure
and catalog vibrations
from the roughing pump.
�
Measure
and catalog sound from
the roughing pump.
� Takes up minimal lab space and is aesthetically pleasing.
�
Notify
to user by email/text
message/automated if pump
parameters go outside the known region of operation.
�
If
no user response to
notification after 1hr pump will power
down.
� Ability for users to remotely power down the pump.
�
Notify the user if the pump parameters go outside the known region of operation.